| Standard Waveforms |
| Sine, triangle, square, pulse, ramp, sinc (sine(x)/x), Guassian pulse, exponential fall, rising pulse, noise, DC |
| Frequency Range | Waveform dependent |
| Source | Internal synthesizer |
| Sine |
| Frequency Range | 100 µHz to 25 MHz (GX1200) 100 µHz to 50 MHz (GX1201) |
| Band Flatness | 5% to 10 MHz; 20% to 50 MHz (GX1200) 5% to 10 MHz; 20% to 25 MHz (GX1201) |
| Programmable Parameters | Start phase, 0° to 360° C |
| Harmonics and Non-Related Spurious | at 5 V pk-pk < -55 dBc for carroer frequencies 1 MHz < -45 dBc for carroer frequencies 5 MHz < -35 dBc for carroer frequencies 10 MHz < -28 dBc for carroer frequencies 25 MHz (GX1200) < -22 dBc for carroer frequencies 50 MHz (GX1201) |
| THD | 0.05% to 100 KHz |
| Triangle |
| Frequency Range | 100 µHz to 12.5 MHz |
| Start Phase | 0° to 360° C |
| Square |
| Frequency Range | 100 µHz to 25 MHz (GX1200) 100 µHz to 50 MHz (GX1201) |
| Duty Cycle | 1% to 99% |
| Rise/Fall Time | <10 ns |
| Aberration | <5% ±10 mV |
| Pulse |
| Frequency Range | 100 µHz to 6.25 MHz |
| Adjustable Parameters | Delay, Rise Time, High Time, and Fall Time: 0% to 99% of period |
| Rise/Fall Time | <10 ns |
| Aberration | <5% ±10 mV |
| Ramp |
| Frequency Range | 100 µHz to 12.5 MHz |
| Adjustable Parameters | Delay, Rise Time, and Fall Time: 0% to 99% of period |
| SINC (SINE (X)/X) |
| Frequency Range | 100 µHz to 3.125 MHz |
| “0” Crossing | 4 to 100 cycles |
| Gaussian Pulse |
| Frequency Range | 100 µHz to 3.125 MHz |
| Time Constant | 10 to 200 |
| Exponential Rise/Fall |
| Frequency Range | 100 µHz to 6.25 MHz |
| Time Constant | 10 to 200 |
| Noise Bandwidth | 25 MHz |
| DC |
| Range | 100% to 100% of amplitude |
| Arbitrary Waveforms |
| Range | 100% to 100% of amplitude |
| Vertical Resolution | 14 bits (16, 384) |
| Waveform Memory | 2 Meg points per standard |
| Download Rate | 5 Meg points per second |
| Memory Segmentation |
| Number of Segments | 1 to 4,096 |
| Min. Segment Size | 16 points |
| Memory Interleave | 4 (Trace lengths must be multiples 4) |
| Sequenced Arbitrary Waveforms |
| Permits division of the memory bank into smaller segements. Segments may be linked and repeated in user-selectable fashion to generate extremely long waveforms. |
Automatic Sequence Advance | No triggers required to step from one segment to the next. Sequence is repeated continuously through a preprogrammed sequence list (table) |
Stepped Sequence Advance | Current segment is sampled continuously, external trigger advances to next programmed segment. Control input is TRIG IN connector. |
Single Sequence Advance | Current segment is sampled to the end of the segment including repeats and idles there. Next trigger advances to next segment. Control input is TRIG IN connector. |
| Mixed Sequence Advance | Each step of a sequence can be programmed to advance a) automatically (automatic Sequence Advance), or b) with a trigger (Stepped Sequence advance) |
| Advance Source | External, internal, or soft trigger |
| Sequencer Steps | From 1 to 4,096 |
| Segment Loops | From 1 to 128 K |
| Common Characteristics |
| Main Output |
| Connector | Front panel BNC |
| Stand-By | Output off or normal |
| Impedance | 50 Ω, +-/1% |
| Protection | Protected against shorts to case ground |
| Frequency |
| Resolution | 10 digits limited by 1 S/s |
| Accuracy | 1 ppm |
| Stability | 1 ppm |
| Reference |
| Standard | 0.0001% (1 ppm TCXO) initial tolerance over a 19° C to 29° C temperature range; 1 ppm/ below 19° C and above 29° C; 1 ppm/year aging rate |
| External | 10 MHz TTL, 50% +/-2% duty cycle |
| Amplitude |
| Range | 80 mV to 8Vpk-pk into 50 O; double into open circuit |
| Resolution | 4 digits |
| Accuracy | 800 mV to 8 V pk-pk ±(1% + 10 mV) 80 mV to 799.9 mV pk-pk ±(1% + 1 mV) |
| Offset |
| Range | 0 to ±3.6 V, amplitude dependent |
| Resolution | 2.2 mV |
Accuracy 4 V Window 400 Mv Window | ±(1% of reading + 1% of amplitude + 2 mV) ±(1% of reading + 1% of amplitude + 200V) |
| Filters | 50 MHz Elliptic 25 MHz Elliptic |
| Sync/Marker Output |
| Provides dual functionality. Sync, which is sychronous with the output waveform, and Marker in FM and sweep modes. |
| Outputs | Front panel SYNC output BNC PXI backplane TTLTrig 0 through 7 programmable |
| Validators | BIT, LCOM |
| Level | >2 V into 50 Ω, 4 V nominal into 10 KΩ |
| Protection | Temporary short case to ground |
| Position | Point 0 to n, programmable with four-point resoltion |
Width Range Resolution | 4 to 100,000 waveform points Programmable with four points resolution |
| Source | Main output |
| Sinewave Output |
| An output that is directly derived from the sample clock generator and has an output frequency equal to the programmed sample clock frequency, including modulated waveform such as FM, sweep, and FSK. |
| Connector | Front panel SMB |
| Impedance | 5 Ω, ±1% |
| Level | 1 V into 50 Ω |
| Protection | Temporary short case to ground |
| Source | Sample clock frequency |
Frequency Range And Resolution | Same as Sample clock |
| Flatness | -3 dB at 100 MHz |
| THD | 0.3% to 100 KHz |
Harmonics and Non-Related Spurious | -55 dBc to 1 MHz -45 dBc to 10 MHz -35 dBc < 50 MHz (GX1200); -35 dBc < 1000 MHz (GX1201) |
| Inputs |
| Trig Input |
| Connectors | Front panel BNC |
| Impedance | 10 KΩ, ±5% |
| Threshold Level | TTL |
| Damage Level | ± 10 V |
| Min Pulse Width | 20 ns |
| Slope | Positive or negative going edge |
| 10 MHz Reference Input |
| Connector | Front panel SMB |
| Impedance | 10 KΩ, ±5% |
| Threshold Level | TTL |
| Damage Level | ± 10 V |
| Duty Cycle | 50%, ±5% |
| Modulation |
Waveform Modulation | Sine, triangle, square, pulse, ramp, sinc (sine(x)x), Gaussian pulse, exponential fall, rising pulse, noise, DC, arb |
| Source | Internal |
| Resolution | 10 digits |
| Accuracy | 0.1% |
Modulating Frequency Distortion | <0.1% |
| Deviation Range | 100 mS/s to 50 MS/s (GX1200) 100 mS/s to 100 MS/s (GX1201) |
| Advanced Mode | Automatic, triggered, gated, or software command |
Marker Output And Level Position | Same as SYNC output Fixed at carrier frequency |
FM - Built-In Standard Waveforms Sample clock can be frequency modulated by internal waveforms that are resident in internal memory (fixed waveforms) |
| Carrier Waveforms | Sine, square, triangle, and ramp |
Modulation Frequency Range | 1 mHz to 100 KHz |
FM - Downloadable Arbitrary Waveforms Sample clock can be frequency modulated by arbitrary waveforms that are downloaded by the user |
| Modulation Source | User waveform, any shape, 10 to 20,000 waveform points |
Modulation Sample Clock Range | 1 MS/s to 2 MS/s |
Waveform Download Rate | 5 Meg points per second |
| FSK |
| Current segment is sampled continuously. TTL low level programs carrier sample clock, TTL high level programs shifted sample clock frequency. Sample clock changes coherently between frequencies. FSK operates on arbitrary waveforms only. |
Carrier Sample Clock Range
| 100 mS/s to 50 MS/s (GX1200) 100 mS/s to 100 MS/s (GX1201) |
| Source | External, front panel, trigger input BNC |
| Frequency Range | From 10 MHz to DC |
| FSK Delay | Minimum one waveform cycle + 50 ns |
| Ramped FSK |
| Same as FSK except carrier sample clock ramps to shifted frequency at a rate defined by the ramp time parameter. TTL low level programs carrier sample clock, TTL high level programs shifted frequency. |
| Ramp Time Range | 10 s to 1s, 3 digits, ±.01% |
| Sweep |
| Sample clock sweeps continuously from start to stop at a rate defined by the sweep time. More complex sweep modes can be generated using the FM mode in conjunction with the FM composer progrm. |
| Carrier Waveforms | Sine, square, triangle, ramp, and arb |
| Type | Linear or logarithmic |
| Direction | Up or down depending on the start and stop settings |
| Range | 100 mS/s to 50 MS/s (GX1200) 100 mS/s to 100 MS/s (GX1201) |
| Time | 1 ms to 1,000 s, 7 digits, ±.01% |
Trigger Advanced Mode | Automatic, triggered, gated, or software command marker output |
| Level | Same as SYNC ouput |
| Position | Programmable for selected frequency |
| Triggering Characteristics |
| Trigger Sources |
| External |
| Connector | Front panel BNC |
| Level | TTL |
| Slope Sensitivity | Positive or negative, programmable |
| Frequency | 5 MHz to DC |
| Impedance | 10 kΩ, DC coupled |
| Internal |
| Range | 100 mHz to 2 MHz |
| Resolution | 7 digits |
| Accuracy | 0.01% |
| Software | SCPI command |
| Backplane | TTLTrig0 through TTLTrig7; STAR |
| Trigger Start Phase |
| Description | Waveform starts from point n and completes at point n-1. |
| Range | 0 to 1,999,999 waveform points |
| Resolution | 4 points |
| Start/Stop Control (Breakpoint) |
| Range | 0 to 1,999,999 waveform points (2 Meg) |
| Source | External (rear panel trigger input BNC) or software command |
| Resolution | 4 points |
| Breakpoint Error | ±4 points |
| System Delay |
Trigger To Waveform Output | One Sample Clock + 120 ns |
| Gated Mode | External signal enables generator. First output cycle synchronous with the activeslope of the triggering signal. Last cycle of output waveform always completed. |
| Burst |
| Waveforms | Sine, triangle, square, pulse, ramp, sinc (sine (x)/x), Gaussian pulse, exponential fall, rising pulse, noise, DC, and arb |
Number Of Cycles Per Burst | 1 to 128 k |
| Trigger Source | Manual, external, or internal- |
| Multiple Instrument Synchronization |
| Multiple instruments can be connected together and synchronized to provide multichannel synchronization. |
| Sample Clock |
| Source | From master card to slave boards through the local bus |
| Range And Resolution | Same as Sample Clock (GX1200) Same as Sample Clock, but limited to 80 MS/s (GX1201) |
| Initial Skew | <20 ns to the first master; 20 ns cumulative to additional slaves |
Phase Offset Between Instruments | Programmable from 0 to n points |
| General |
Power Requirements | 10 W (Max.) |
Current Consumption | +5 V @ 30 mA +12 V @ 200 mA -12 V @ 200 mA +3.3 V @ 1.4 A |
| Weight | Approx. 0.5 kg |
| Size | 3U |
Temperature Operating Storage | 0° C to 40° C -20° C to 70° C |
Humidity (Non-Condensing) | 11° C to 30° C: 85% 31° C to 40° C: 75% |
Workmanship Standards | Conform to IPC-A-610D |
| Reliability | MTBF per MIL-HDBK-217E, 25° C, ground benign |
| Safety | Designed to meet IEC 1010-1, UL 3111-1, and CSA 22.2#1010 |
| CE Labeled | Yes |
Supplied Accessories | CD containing Manual ArbConnection. ArbDetector, and devloper libraries |
| GPIB Interface |
| Compatibility | Conforms to IEEE488.2 and is compatible with SCPI (Standard Commands for Programmable Instruments). |
| Subsets | SH1, AH1, T6, L4, SR1, RL1 PPO, DC1, DT1, CO, E2. |
|